4Q-MEPS
Motor Efficiency Certification Platform
Turn-key IEC 60034-2-1 motor efficiency certification labs from 0.12 kW to 375 kW.
01 – THE CHALLENGE
Why Motor Efficiency Labs Fail
Establishing a motor efficiency certification laboratory has traditionally meant procuring equipment from multiple vendors. Each integration point between instruments is a source of calibration error, data integrity risk, and support complexity.
When measurement uncertainty compounds across instruments, an error of 0.8% or more is enough to misclassify an IE3 motor as IE2 – and no amount of post-test data correction can recover a result built on miscalibrated inputs.
Common Integration Problems
  • ■Timing mismatch between input and output power measurements
  • ■No unified calibration traceability across instruments
  • ■VFD switching harmonics corrupting input power measurement
  • ■Manual test sequencing – operator-dependent variation
  • ■No standardized report format for certification bodies
  • ■Separate service contracts – weeks of downtime during disputes
The 4Q-MEPS Approach
  • ■Single factory-calibrated system – all instruments matched
  • ■Synchronized acquisition – input and output measured simultaneously
  • ■Sinus filter eliminates VFD harmonics from the measurement loop
  • ■Automated 12-step IEC 60034-2-1 sequence – no manual steps
  • ■Structured PDF test reports compliant with IEC 60034-2-1 reporting requirements
  • ■One service contract – one call resolves everything
The 4Q-MEPS platform eliminates integration risk by delivering all measurement, control, and reporting functions from a single, pre-calibrated, factory-tested system.
02 – SYSTEM ARCHITECTURE
Four-Quadrant Regenerative Design
Regenerative Common DC Bus
The 4Q-MEPS uses a four-quadrant regenerative dynamometer coupled to the motor under test via a precision torque transducer. The load machine and motor under test share a common DC bus — energy recovered circulates internally, and only losses are drawn from the supply grid.
Input power is supplied through an auto-transformer for voltage setting and a sinus filter that removes VFD switching harmonics from the measurement loop. The motor under test sees a clean sinusoidal supply — the same conditions as an installed drive.
  • ■VFD + sinus filter + Auto Transformer
  • ■Common DC bus
  • ■Auto-transformer: all voltage ratings 20-690 V
  • ■Torque transducer: NIST-traceable, 0.1% FS accuracy
03 – TEST AUTOMATION
IEC 60034-2-1 Method B – Test Procedure
The 4Q-MEPS controller executes the full IEC 60034-2-1 Method B test procedure automatically. The operator enters motor nameplate data – the system manages the complete sequence across four phases: test preparation, test execution, loss calculations, and efficiency output.
01 – Test Preparation02 – Test Execution03 – Loss Calculations04 – Efficiency Output
Stator winding resistance at ambient temperatureStator winding resistance measurementConstant losses (Cl. 8.2.2.3.1) – Friction & windage losses (8.2.2.3.2)Calculate efficiency (Cl. 8.2.2.1)
Remove accessible drive-end seal(s) without dismantlingNo-load test (Cl. 6.4.2.3)Iron losses (Cl. 8.2.2.3.3)Calculate total losses (Cl. 8.2.2.2)
Rated load temperature test (Cl. 6.4.4.1)Stator winding resistance measurementLoad losses (Cl. 8.2.2.4.1) – Stator & rotor winding losses incl. temp. correctionGamma value > 0.95 and Intercept B < 50% PLL
Stator winding resistance measurementLoad curve test with torque measurement (Cl. 6.4.4.2 and 6.4.5.3)Calculate residual losses PLr – Refer equation 21Additional load losses PLL to Cl. 8.2.2.5.1
Operator enters motor nameplate data. The system runs the full IEC 60034-2-1 Method B procedure, confirms thermal stability, captures all measurements across every clause, and generates the certification report automatically.
04 – PLL POWER FACTOR CONTROL
Eliminating Reactive Power Errors
Why PLL Matters in Motor Certification
Induction motor efficiency testing at low loads produces power factors below 0.5. At such low power factors, a phase angle error of just 0.5 degrees causes a disproportionately large active power measurement error – the leading source of uncertainty in input power at partial load.
The 4Q-MEPS integrates a Phase-Locked Loop auto-phase correction system that continuously monitors the phase angle between voltage and current, applying real-time corrections regardless of load condition.
  • ■Minimum power factor enforced: gamma >= 0.95
  • ■Target power factor maintained: gamma >= 0.98
  • ■Fast PLL update rate to track all transients
  • ■Auto-alert if phase correction limit is approached
PLL correction reduces input power measurement uncertainty from the typical 0.5-1.5% of reactive-loaded setups to under 0.15% – mandatory for IE4 and IE5 certification where the margin between efficiency classes is less than 1%.
05 – PRODUCT RANGE
Three System Sizes for Every Lab
MEPS-S

0.12 kW – 7.5 kW
Small motors, fractional kW, pumps and fans.
MEPS-M

7.5 kW – 55 kW
Standard industrial motors – the core MEPS/IE efficiency market.
MEPS-L

55 kW – 375 kW
Large motors for pumps, compressors, HVAC and industrial drives. Multi-cabinet system for high-power certification labs.
06 – STANDARDS AND COMPLIANCE
Accepted by Certification Bodies Worldwide
Every 4Q-MEPS system ships with pre-built test sequences for all supported standards. Select the standard, enter motor nameplate data, and run – no programming required.
StandardScopeRegionIE Class
IEC 60034-2-1AC motors – efficiency measurement methodsGlobalIE1 – IE5
IEC 60034-30-1AC motors – efficiency classes definitionGlobalIE1 – IE5
NEMA MG-1US motors – efficiency test and classificationUSANEMA Nom.
GB 18613Chinese mandatory motor efficiency standardChinaIE2 – IE4
IS 12615Indian motors – energy performance standardsIndiaIE2 – IE4
Every 4Q-MEPS system is delivered with a full documented measurement uncertainty budget covering every instrument in the chain. Typical measurement accuracy for motor efficiency: better than +-0.2% at 100% load.
Build Your MEPS Certification Lab with 4Q Systems
Turn-key delivery, installation, and commissioning – ready to certify in 3 days.
Related 4Q Testing Platform
The MEPS Certification Labs application platform is based on the 4Q-MEPS motor efficiency certification system, designed for IEC 60034-2-1 validation, efficiency mapping, temperature rise testing, and laboratory certification environments.


Explore 4Q-MEPS Platform →